JY6611 Transformer Oil Breakdown Voltage (BDV) Tester Demo Video:
JY6611 transformer oil breakdown voltage tester (oil dielectric strength tester,oil BDV tester) adopts completely new electromagnetic compatibility concept(EMC) which designed to avoid tester crash accident during testing, and also can still work well in high magnetic field.
JY6611 transformer oil breakdown voltage tester (oil dielectric strength tester,oil BDV tester) use advanced voltage breakdown control technology which will keep breakdown energy in a very low level,that will avoid pollution to oil sample during testing and make sure the test result are accurate and reliable.
JY6611 also use advanced voltage breakdown control technology which will keep breakdown energy in a very low level, that will avoid pollution to oil sample during testing and make sure the test result are accurate and reliable.
The unique advantage of this tester lies in the built-in electronic boost system, the voltage regulation power supply adopts an electronic inverter sine wave generator, the voltage output is accurate, the waveform quality is high, and it is not affected by the grid voltage fluctuation and waveform distortion, making the test data more accurate and efficient. . The built-in advanced insulating material and heat dissipation system make the built-in step-up transformer stable and reliable, which can make the tester withstand long-term high-voltage breakdown tests, protect the safety of operators, and prolong the service life of the tester.
JY6611 has a variety of protection devices which will make sure the safety of operator and tester itself in accident like inferior oil breakdown voltage and empty cup breakdown.
JY6611 transformer oil breakdown voltage tester (oil dielectric strength tester,oil BDV tester) select series NEC Microprocessor to make sure high testing accuracy
Oil test cup made from advanced material which is high intensity with long working life.
Thermosensitive printer provide fast and clear tested value printing.
Mashroom shape and ball shape electrodes as standby application:
The gap between electrodes is easily regulated with a use of pass-no pass gauge, which comes in pack.
TECHNOLOGY SPECIFICATION:
Type |
JY6611 oil break down voltage(BDV) tester |
Output voltage |
0~80kV/100kV |
Minimum resolution |
0.1kV |
Accuracy |
Reading±0.2kV |
Voltage raise speed |
1 .0/2.0/3 .0kV/s for option |
Switch-Off time on breakdown, ms |
≤1ms |
Experimental times |
1~6 for option |
Programmed test standard |
IEC60156/ASTM D877/ASTM D1816/VDE0370 |
Memory for result |
500 group |
Volume of the oil cup |
400ml/200ml |
Gap between the poles |
2.5mm (Electrode gap is adjustable) |
Temperature in operation |
0~40℃ |
Relative humidity |
≤80%RH, No forming dew |
Power supply in operation |
AC220V±10%, 50Hz±1% |
Volume / Weight |
Length 385mm Width 300mm Height 360mm / 22kgs |
CALIBRATION CERTIFICATE FROM ILAC.MRA(The world's most stringent calibration testing agency)
All KRI tests will adopt international express to reach any corner of the world as fast as 72 hours